Compressor NebulizerNE-C801INSTRUCTION MANUALTMCaution: Federal (USA) law restricts this device to sale by or on the order of a physician or licensed
10KNOW YOUR UNITComponents:Instruction Manual Storage BagAir FiltersPackage of 5Compressor NebulizerNE-C801INSTRUCTION MANUALCaution: Federal (USA) l
11WARNING• Clean and disinfect the nebulizer kit, mouthpiece, and optional nosepiece or optional masks before using them for the first time after pur
123 Rotatetheinhalationtopcounterclockwise(A)andlift(B)to remove the inhalation top from the medication cup. 4 Add the correct amount of
13PREPARING THE NEBULIZER FOR USE7 Attach the desired inhalation accessory. USING THE MOUTHPIECE Place the cap on the medication cup over the inha
141 Twist the air tube plug slightly and push it firmly into the air tube connector on the upper side of the compressor (A). 2 Twist the air tube
15WARNING• Do not use the device where the device may be exposed to flammable gas or vapors.• Do not use or store the device in humid locations, suc
16CAUTION• Do not block the air filter cover.• Limit use to 20 minutes at a time, and allow a 40 minutes interval before using the device again.• Wh
175 Turn the power switch to the off ( ) position. The compressor turns off and nebulization stops.6 Disconnect the air tube from the nebulizer ki
18Following cleaning instructions after each use will prevent any remaining medication in the bottle from drying resulting in the device not nebulizin
19DAILY DISINFECTINGDisinfect the nebulizer kit and mouthpiece, or optional masks or optional nosepiece after the last treatment of the day.To disinfe
2CONTENTSThe illustrations shown in this instruction manual are image drawings and may differ slightly from the appearance of the actual product. Befo
20If the air filter has changed color or has been used on average for more than 60 days, replace it with a new one. If water or medication has spilled
21To keep your device in the best condition and protect the unit from damage, follow these directions:CAUTION• Do not subject the compressor or any o
22SOLUTIONCAUSEPROBLEMNo power to unit when the power switch is on. Turn the power switch off.Plug the AC adapter into an electrical outlet. Turn the
23Your OMRON NE-C801 Compressor Nebulizer, excluding the nebulizer kit, inhalation accessories, air tube and air filters, is warranted to be free from
24NOTE:POTENTIALFORRADIO/TELEVISIONINTERFERENCE(forU.S.A.only)ThisproducthasbeentestedandfoundtocomplywiththelimitsforaClassBdi
25Model: NE-C801 (NE-C801S-Z)Type: Compressor nebulizerRating (AC adapter): 100-240V~350mA,50/60HzRating (Compressor Nebulizer): 12V0.8ANebu
26Particle Specifications A series of aerosol performance tests were performed using an 8 stage cascade impactor at a sampling flow r
27Important information regarding Electro Magnetic Compatibility (EMC)With the increased number of electronic devices such as PC’s and mobile (cellula
28Electromagnetic immunity IEC60601-1-2Immunity testIEC 60601Test levelCompliance levelElectromagnetic environment – guidanceElectrostatic discharge (
29Electromagnetic immunity IEC60601-1-2Immunity testIEC 60601Test levelCompliance levelElectromagnetic environment – guidanceConducted RFIEC 61000-4-6
3Thank you for purchasing the OMRON NE-C801 Compressor Nebulizer.Fill in for future reference.DATE PURCHASED: __________________SERIAL NUMBER: _______
30Recommended separation distance between portable and mobile RF communications equipment and the NE-C801The NE-C801 is intended for use in an electro
NE-C801S-Z_A_M_E_SP_041311.indd 31 4/13/11 10:20:59 AM
active forever.comwww.ActiveForever.com
4The NE-C801 Compressor Nebulizer System is intended to provide air to the pneumatic nebulizer in order to aerosolize medications for inhalation by th
5To ensure the correct use of the product, basic safety measures should always be followed including the warnings and cautions listed in this instruct
6OPERATING THE DEVICE (continued) Do not operate where oxygen is being administered in a closed environment (such as an oxygen tent). Provide close
7OPERATING THE DEVICE (continued) Dispose of the device, components and optional accessories according to applicable local regulations. Unlawful disp
8SAFETY INFORMATIONRISK OF ELECTRICAL SHOCK (continued) Do not overload power outlets. Plug the AC adapter into the appropriate voltage outlet.
9Air Tube (PVC, 100 cm)Nebulizer Kit HolderPower SwitchNebulizer KitCapInhalation Air InletInhalation TopBaffleMedication CupNozzleAir Tube ConnectorM